Product/Service

HUMOR10 Humidity Generator

Source: JLC International, Inc.
Humidity measurement and its control play an important role in everyone's life, including saving energy costs and also to ensure optimum climate conditions in production processes
HUMOR10

Humidity Generator

Main Features
  • Highest Accuracy
  • Compact Housing
  • Large Measuring Chamber
  • Traceable to Internal Standards
  • Available with Certificate
  • Easy Handling
Humidity measurement and its control play an important role in everyone's life, including saving energy costs and also to ensure optimum climate conditions in production processes.

This requires a large assortment of different humidity measuring instruments that are available on the market.  They all differ in regard to housing design and in their measuring accuracies and long term stability.

E+E ELEKTRONIK, as one of the leading manufacturers of humidity transmitters is frequently confronted with this problem.   As a result of our extensive know-how in the field of humidity measurement has enabled us to develop a compact and easy to use humidity calibration and data acquisition system of high accuracy.

The measuring method of the portable HUMOR10 is based on a "two-pressure generator", the same operational function as is used by the national Bureau of Standards for very accurate humidity calibrations.  Based on this principle the generator is reliable and traceable to international standards.  By request it is available with an official test-certificate.

Systematic errors are adjustable on the existing serial interface as a calibration function.  This leads to an outstanding accuracy of ± 0.3 % RH in the measuring range between 10 and 85 % RH.  With reduced accuracy it is also possible to use the HUMOR10 up to 95 % RH.

The only equipment required is a mains supply of 230 (110) volts AC and a filtered compressed air line, which should be oil free.  Air conditioned room facilities are not necessary.

The Generator enables the customer to install and calibrate either cylindrical humidity sensor probes each with a maximum diameter up to 16 mm or a cubic measuring unit with maximum dimensions of 80 x 80 x 20 mm.

Only the inlet pressure of the system is required to be adjusted with a rotary control button.  The evaluation of the relative humidity values is carried out by an integrated high performance microprocessor.

The calibration system is equipped with an online display for reading the corresponding measurements.  The stabilization time of the HUMOR10 is less than one minute.

A power supply of 24 VDC is available for the transmitters.
  
 

Accessories and Spare
For the reliable operation of E+E transmitter series, several accessories and spare parts are available:
  • Calibration set for rechecking and adjusting the humidity calibration
  • Radiation shield used with our transmitters for outdoor applications
  • Dust filter caps designed for different applications
  • Installation flanges for duct mounting
  • Programmable display module for (almost) any type of application - Measuring / Displaying / Controlling / Supervising / Counting
      
Technical Data
Principal function two pressure-reactor
Working Range 10-95%RH
Accuracy 10 - 70% RH       ± 0.3% RH
70 - 85% RH      ± 0.5% RH
85 - 95% RH      ± 1.0% RH
Setting Time < 1 min.
Stability < 0.2% RH/h
Gas Flow 3 1/min
for RH > 85% RH the gas flow is reduced to 1.5 1/min at 95% RH
Supply 230VAC (110AC)
compressed air, filtered and free of oil or nitrogen (N2) with 10 < p < 12 bar
Integrated power supply 24VDC, 150mA
Display two line Display for RH and commentary
Serial bus type RS 232 C
Dimensions 400 x 260 x 240mm
Weight about 23 kg

JLC International, Inc., 55 Hawk Road, Suite 175, Warminster, PA 18974. Tel: 1-800-599-4732; Fax: 1-215-444-9219.